Testing high level out, Measuring peak voltage, current, and overshoot – Fluke Biomedical QED 6 User Manual
Page 46

QED 6
Users Guide
2-20
Testing High Level Out
All waveforms available through the ECG jacks are simultaneously output
through the High Level jacks. This scheme offers the user a 1-volt peak signal
for testing purposes.
To test a High Level signal, use an oscilloscope and a scope probe to measure
the output waveform on the high level output. Refer to Generating Test
Waveforms, above, for procedure.
Note
Use a scope probe to guarantee signal integrity.
Measuring Peak Voltage, Current, and
Overshoot
To measure the peak voltage and current of the defibrillator pulse:
1. Power up the defibrillator to be tested and select the energy output,
following the manufacturer’s instructions.
2. Power up the Analyzer.
Main Menu 1
displays:
fcf002.eps
3. Press the
PEAK
softkey to access the
--- V --- A
(current and voltage)
menu:
fcf025.eps
Pressing the
Peak / OVER
softkey toggles the measurement between
Peak
current and voltage and
Over
current and voltage.