Operation pacemaker (non-invasive) testing – Fluke Biomedical QED 6 User Manual
Page 49

Operation
Pacemaker
(Non-Invasive)
Testing
2
2-23
QED-6
DEFIBRILLATOR ANALYZER
STERNUM (-)
BATTERY / 9VDC
RS-232 / 10101
ECG
V
C
LL
F
LA
L
RL
N
RA
R
DEFIB
APEX (+)
PACER
1000 j
100 j
SOFTKEYS
O
u
tp
u
t
Pacer
Adapters
Pacer
ECG Leads
RA
LA
LL
fcf015.eps
Figure 2-9. Connecting Pacemaker Output to Analyzer
Note
The Pacer can be in either demand or non-demand mode.
5. Press the
MEAS
softkey to display the following:
fcf031.eps
Three hyphens (---) indicate that no pacer pulses were received.
6. Set the pacer at various current and heart rate settings. The results are
displayed. Press the
softkey for a hard copy.
See also other documents in the category Fluke Biomedical Equipment:
- 7000DP Impulse Supplement (12 pages)
- 7000DP Impulse Getting Started (36 pages)
- 7000DP Impulse (92 pages)
- 06-526-2200 (26 pages)
- 07-417 (16 pages)
- 07-443 (14 pages)
- 07-451 (8 pages)
- 07-487 (12 pages)
- 07-453 (24 pages)
- 07-555 (8 pages)
- 07-553 (10 pages)
- 07-605-7777 (8 pages)
- 07-444 (46 pages)
- 07-618 (10 pages)
- 07-600 (6 pages)
- 07-591 (8 pages)
- SigmaPace 1000 (154 pages)
- 07-653 (8 pages)
- 07-633 (20 pages)
- 07-649 (14 pages)
- 07-661-7662 (12 pages)
- 07-645 (14 pages)
- 10100AT (80 pages)
- 07-644 (10 pages)
- 18-203 (4 pages)
- 07-621 (12 pages)
- 07-647 (12 pages)
- 18-207 (8 pages)
- 18-216-1000 (10 pages)
- 18-220 (10 pages)
- 18-228 (8 pages)
- 18-223 (18 pages)
- 18-229-1313 (6 pages)
- 18-250 (8 pages)
- 18-252 (6 pages)
- 18-303 (6 pages)
- 35035 (19 pages)
- 6000-528 (18 pages)
- 35080B (70 pages)
- 6000-529 (16 pages)
- 35080M (64 pages)
- 6000-530B (16 pages)
- 57-436 (12 pages)
- 57-440 (12 pages)
- 57-402 (32 pages)