Ecg performance, Pacer – Fluke Biomedical QED 6 User Manual
Page 56

QED 6
Users Guide
2-30
ECG Performance
1. Press
STEP
to advance to the next check item,
ECG/Perf
.
fcf043.eps
2. Press
+
or
-
to advance to the next waveform.
3. Press
SEL
to program / deprogram a waveform. An * indicates that the
item is programmed.
Pacer
1. Press
STEP
to advance to the next check item,
Pacer
.
fcf044.eps
2. Press the
SEL
softkey to toggle between
Yes
and
No
.
3. Press
STEP
to return to the program menu.
fcf045.eps
4. Press the
esc
softkey. You are asked, “
Save Changes?
”
5. Press the
YES
softkey to save the program or the
NO
softkey to return
without saving; both return to the
Autosequences
menu:
fcf046.eps
6. Press the
esc
softkey to return to
Main Menu 2
.
See also other documents in the category Fluke Biomedical Equipment:
- 7000DP Impulse Supplement (12 pages)
- 7000DP Impulse Getting Started (36 pages)
- 7000DP Impulse (92 pages)
- 06-526-2200 (26 pages)
- 07-417 (16 pages)
- 07-443 (14 pages)
- 07-451 (8 pages)
- 07-487 (12 pages)
- 07-453 (24 pages)
- 07-555 (8 pages)
- 07-553 (10 pages)
- 07-605-7777 (8 pages)
- 07-444 (46 pages)
- 07-618 (10 pages)
- 07-600 (6 pages)
- 07-591 (8 pages)
- SigmaPace 1000 (154 pages)
- 07-653 (8 pages)
- 07-633 (20 pages)
- 07-649 (14 pages)
- 07-661-7662 (12 pages)
- 07-645 (14 pages)
- 10100AT (80 pages)
- 07-644 (10 pages)
- 18-203 (4 pages)
- 07-621 (12 pages)
- 07-647 (12 pages)
- 18-207 (8 pages)
- 18-216-1000 (10 pages)
- 18-220 (10 pages)
- 18-228 (8 pages)
- 18-223 (18 pages)
- 18-229-1313 (6 pages)
- 18-250 (8 pages)
- 18-252 (6 pages)
- 18-303 (6 pages)
- 35035 (19 pages)
- 6000-528 (18 pages)
- 35080B (70 pages)
- 6000-529 (16 pages)
- 35080M (64 pages)
- 6000-530B (16 pages)
- 57-436 (12 pages)
- 57-440 (12 pages)
- 57-402 (32 pages)