Defib waveform playback, Arrhythmias, Scope outputs – Fluke Biomedical QED 6 User Manual
Page 24

QED 6
Users Guide
1-12
Defib Waveform Playback
Time Base Expansion .............................................100:1 @ 25 mm/sec paper speed;
each division equals 40 ms
Amplitude Scaling
Lead II (RA-LL)
1000 J Range ..................................................1 mV = 3000 V
100 J Range ....................................................1 mV = 900 V
ECG Output
1000 J Range ..........................................................0.5 V = 3000 V
100 J Range ............................................................0.5 V = 900 V
Arrhythmias
Afib, Vfib, Atach, Vtach, Aflutter, RUN, PVC, R on T, Idioventricular
Rate Accuracy .........................................................±1 %
Amplitude.................................................................Fixed at 1 mV Lead II
(RA-LL)
Fixed at 1.1 mV (Apex-Sternum)
Amplitude Accuracy.................................................±2 (RA-LL) ±10 % (Apex-Sternum)
Scope Outputs
ECG Hi-Level...........................................................Fixed at 1 V
Accuracy: ±2 %
Defib Output ............................................................Real Time
Pacer Range............................................................1 V = 3.11 V
1000 J Range ..........................................................1 V = 1450 V
100 J Range ............................................................1 V = 440 V
Amplitude Accuracy.................................................±2 % of scale