Front panel – Fluke Biomedical QED 6 User Manual
Page 20

QED 6
Users Guide
1-8
Front Panel
The front panel of the Analyzer includes the elements described in Table 1-2.
Table 1-2. Front Panel Elements
Number Element
Function
A
Universal ECG
jacks
Utilize AHA and International color coding,
allowing for waveform output to monitor/recorder.
B
High Level
ECG Banana
jack
Provides 1 volt peak output of the selected
waveform.
C
Common
Banana jack
Provides ground for the “High Level ECG” and
“Defib Scope Out” jacks.
D
Defib Scope
Out Banana
jack
Provides pulse output to an oscilloscope.
E
Range switch
Allows for defibrillator settings from 0 to 1000
joules (high), for power below 0-100 joules (low)
for increased accuracy, and a PACER range
setting for pacer output measurements.
F
Power switch
Enables the Analyzer (I = ON, O = OFF).
G
LCD display:
24 characters x
2 lines.
The upper line of the LCD display provides
messages and test results, while the bottom line
displays menu choices.
H
Five softkeys
Used to select the desired function highlighted on
the lower line of the display.
I
Two nickel-
plated
Defibrillator
Paddle Plates
Available for defibrillator paddle contact. All
waveforms are present at the paddle plates
simultaneously with the ECG jacks.