Fluke Biomedical QED 6 User Manual
Page 43

Operation
Measuring
Synchronization
2
2-17
Sync time measurements are performed as shown in Figure 2-8, below:
R-
W
a
v
e Time
R
QRS COMPLEX
90 BPM
N
SR
Q
Defi
b
P
u
lse
Peak
Q-
W
a
v
e Time
DEFIB PULSE
fcf019.eps
Figure 2-8. Sync Time Measurements
5. Set the range switch appropriately, as follows:
•
Select the 1000 joule high range for defibrillator outputs over 100
joules or for an unknown defibrillator output power.
•
Select the 100 joule low range for outputs under 100 joules.
6. Place the defibrillator in synchronous mode.
7. Simultaneously press both defibrillator paddles to the contact plates of the
Analyzer.
8. Initiate a discharge from the defibrillator.
9. Press the
ENRG
softkey to view the energy readings on the display, as
shown below.
See also other documents in the category Fluke Biomedical Equipment:
- 7000DP Impulse Supplement (12 pages)
- 7000DP Impulse Getting Started (36 pages)
- 7000DP Impulse (92 pages)
- 06-526-2200 (26 pages)
- 07-417 (16 pages)
- 07-443 (14 pages)
- 07-451 (8 pages)
- 07-487 (12 pages)
- 07-453 (24 pages)
- 07-555 (8 pages)
- 07-553 (10 pages)
- 07-605-7777 (8 pages)
- 07-444 (46 pages)
- 07-618 (10 pages)
- 07-600 (6 pages)
- 07-591 (8 pages)
- SigmaPace 1000 (154 pages)
- 07-653 (8 pages)
- 07-633 (20 pages)
- 07-649 (14 pages)
- 07-661-7662 (12 pages)
- 07-645 (14 pages)
- 10100AT (80 pages)
- 07-644 (10 pages)
- 18-203 (4 pages)
- 07-621 (12 pages)
- 07-647 (12 pages)
- 18-207 (8 pages)
- 18-216-1000 (10 pages)
- 18-220 (10 pages)
- 18-228 (8 pages)
- 18-223 (18 pages)
- 18-229-1313 (6 pages)
- 18-250 (8 pages)
- 18-252 (6 pages)
- 18-303 (6 pages)
- 35035 (19 pages)
- 6000-528 (18 pages)
- 35080B (70 pages)
- 6000-529 (16 pages)
- 35080M (64 pages)
- 6000-530B (16 pages)
- 57-436 (12 pages)
- 57-440 (12 pages)
- 57-402 (32 pages)