Fluke Biomedical 07-444 User Manual
Page 29
![background image](https://www.manualsdir.com/files/785232/content/doc029.png)
Setup
Step Wedge Type-X-Rite
4
4-15
5. Analyzing data... will be displayed.
6. Ready for scan 2, 3, etc. will be displayed, repeat the scanning technique described above until the
previously selected number of scans has been completed.
7. The previous or default value will be displayed Grad Dev (or CIndx Dev): 0.XX. Edit the entry by
pressing or to increase or decrease the value at the cursor. Press Enter to accept the current
cursor value and advance the cursor to the next digit. When the entry has been completed the
display will advance to the next chart item.
8. The previous or default value will be displayed Speed Dev (or SIndx Dev): 0.XX. Edit the value as
previously described.
9. The previous or default value will be displayed B + F Dev: 0.XX. Edit the value as previously
described.
10. The previous or default value will be displayed Max OD Dev: 0.XX. Edit the value as previously
described. When the entry has been completed Enter New Params is displayed.
11. To exit file, press or to scroll through the available selections until Exit File XX is displayed.
Press ENTER.
12. To review the chart file values, press or to scroll through the available selections until Enter New
Params is displayed. Press ENTER. Continue to press ENTER to review each items baseline and
deviation. See section ENTER NEW PARAMETERS for further description.
Enter New Parameters
This procedure allows the user to manually setup the parameters in a chart file. User entry for each
baseline and deviation value may be entered. This mode may also be used at any time to review the
various values without changing them.
Follow the steps under chart file setup before proceeding.
1. Press or to scroll through the selected file setup functions until Enter New Params is displayed.
Press ENTER.
2. The previous or default value will be displayed Grad Base (or CIndx Base): X.XX. Edit the entry by
pressing or to increase or decrease the value at the cursor. Press Enter to accept the current
cursor value and advance the cursor to the next digit. When the entry has been completed the
display will advance to the next chart item.
3. The previous or default value will be displayed Grad Dev (or Cindx Dev): 0.XX. Edit the value as
previously described.
4. The previous or default value will be displayed Speed Base (or SIndx Base): X.XX. Edit the value
as previously described.
5. The previous or default value will be displayed Speed Dev (or SIndx Dev): 0.XX. Edit the value as
previously described.
6. The previous or default value will be displayed B + F Base: X.XX. Edit the value as previously
described.
7. The previous or default value will be displayed B + F Dev: 0.XX. Edit the value as previously
described.
8. The previous or default value will be displayed max OD Base: X.XX. Edit the value as previously
described.
9. The previous or default value will be displayed Max OD Dev: 0.XX. Edit the value as previously
described. When the entry has been completed Clear Chart File is displayed.
10. To exit file, press or to scroll through the available selections until Exit File XX is displayed.
Press ENTER.