Appendix, A. auto cal buffer tables, B. t80 hart menu – Electro-Chemical Devices (ECD) T80 User Manual
Page 67: A. auto cal buffer tables b. t80 hart menu, Hart
APPENDIX
A.
A
UTO
C
AL
B
UFFER
T
ABLES
°C
pH
pH
pH
0
4.00
7.115
10.32
5
4.00
7.085
10.25
10
4.00
7.06
10.18
15
4.00
7.04
10.12
20
4.00
7.015
10.06
25
4.005
7.00
10.01
30
4.015
6.985
9.97
35
4.025
6.98
9.93
40
4.03
6.975
9.89
45
4.045
6.975
9.86
50
4.06
6.97
9.83
55
4.075
6.97
60
4.085
6.97
65
4.10
6.98
70
4.13
6.99
75
4.14
7.01
80
4.16
7.03
85
4.18
7.05
90
4.21
7.08
B.
T80
HART
M
ENU
T80 Root Menu
1 DEVICE SETUP
DEVICE SETUP
CALIBRATION
1 CALIBRATION
1 AUTO
2 STANDARDIZE
3 MANUAL
4 TEMP
2 BASIC SETUP
CONFIG
1 XMTR
XMTR CONFIG
ANALOG CONFIG
1 ANALOG
1 SCALE
ANALOG SCALE
1 UPPER
2 LOWER
2 PID
PID CONFIG
1 P Term
2 I Term
3 D Term
3 CALIBRATE
CALIBRATE
1 Zero cal
2 Gain cal
3 Save
2 ALARMS
ALARM CONFIG
1 Alrm 1 thresh
2 Alrm 1 hyst
66