Page 45: 2 standardize

4.8.1
S
TANDARDIZE
With the sensor in the process solution, take a grab sample from the process and determine the TDS using a
qualified laboratory method.
Action
Prompt
Double Press any Button
MENU HOME, Hold is OFF
Press HOLD
Hold freezes 4-20 mA Output and locks Alarm Relays during Calibration
Press CAL
Is this a new Sensor? (Yes, erases CAL Log in INFO, NO adds CAL to existing Log)
Press NO
With the sensor in the process solution or calibration standard
Press STAND
Enter the laboratory determined TDS value
Press
▲ ▼ NEXT
xxx.xx ppm (use arrows and NEXT to enter process value)
Press OK
xxx.xx ppm, xxx.x mV, Accept Value?
Press YES
Current value: xx.xx ppm, Desired value: xx.xx ppm, Offset: xx.xx ppm
Press OK
Back to Cal Menu
Press HOME
Hold is ON (Press HOLD to turn off Hold)
Press HOLD
Turn off Hold
Press EXIT
Main Display
4.9
R
ESISTIVITY
S
ENSORS
AUTO Calibration recognizes Air for zero point (Cal 1) and a Meg-Ohm process solution for the span (Cal 2). Use
the actual process solution or a solution near the resistivity of the process solution. The example below uses air
and 15 MΩ. The Cal 1, air calibration value is very stable and need not be done for every calibration.
4.9.1
AUTO
C
AL USING
A
IR AND
M
EG
-O
HM PROCESS WATER
Action
Prompt
Double Press any Button
MENU HOME, Hold is OFF
Press HOLD
Hold freezes 4-20 mA Output and locks Alarm Relays during Calibration
Press CAL
Is this a new Sensor? (Yes erases CAL Log in INFO, NO adds CAL to existing Log)
Press Yes/NO
Remove the Sensor from Solution and dry the front end
To perform zero CAL press CAL 1, to skip zero cal press CAL 2, only if a zero cal was previously done on the sensor
Press AUTO then CAL 1
STABILIZING, 55.000 MΩ, 1.20 V, 55.000 MΩ corrected, Accept Cal?
Press YES
CAL 1 Value 55.000 MΩ, Continue to CAL2? Move sensor to MΩ solution
Press YES
STABILIZING, 9.875 MΩ, 517 mV, 9.875 MΩ corrected, Accept Cal?
Press NO
Enter actual MΩ value of the solution, Accept Cal
Press YES
OFFSET: 55.000 MΩ, 1.20 V , SLOPE: 0.938 (data written to Log)
Press OK
Calibration complete
Press HOME
Hold is ON
Press HOLD
Turn off Hold
Press EXIT
Main Display
4.9.2
S
TANDARDIZE
With the sensor in the process solution, take a grab sample from the process and determine the conductivity
using a qualified laboratory conductivity meter.
Action
Prompt
Double Press any Button
MENU HOME, Hold is OFF
Press HOLD
Hold freezes 4-20 mA Output and locks Alarm Relays during Calibration
Press CAL
Is this a new Sensor? (Yes, erases CAL Log in INFO, NO adds CAL to existing Log)
44