12 silver electrode, 13 sodium electrode, 14 sulfide electrode – Electro-Chemical Devices (ECD) T80 User Manual
Page 65: 15 triton® do80

Temperature Range: 0°-40°C
Pressure Range: 50 psig
8.4.12
S
ILVER
E
LECTRODE
Part #: 2005016 (2005316 SENTINEL)
ISE sensor: solid state AgS membrane
Construction: PEEK body,
Reference Electrode: double porous Teflon junction
Measurement range: 0.1-107,000 ppm
pH Range: 2-14 pH
Temperature Range: 0°-80°C
Pressure Range: 150 psig
8.4.13
S
ODIUM
E
LECTRODE
Part #: 2005031 (2005331 SENTINEL)
ISE sensor: Sodium selective Glass membrane
Construction: PEEK body,
Reference Electrode: double porous Teflon junction
Measurement range: 0.2 - 23,000 ppm
pH Range: 2-14 pH (pH must be 3 units higher than pNa)
Temperature Range: 0°-80°C
Pressure Range: 150 psig
8.4.14
S
ULFIDE
E
LECTRODE
Part #: 2005122 (2005322 SENTINEL)
ISE sensor: solid state AgS membrane
Construction: PEEK body,
Reference Electrode: double porous Teflon junction
Measurement range: 0.01 - 32,000 ppm
pH Range: 11-14 pH
Temperature Range: 0°-80°C
Pressure Range: 150 psig
8.4.15
TRITON®
DO80
Measurement Range:
0 - 20 mg/l (0 - 20 ppm)
0 - 200 % Saturation
0 - 400 mbar (0 - 6 psi)
Pressure Range:
Maximum Pressure 10 bar (159 psi)
Temperature Range:
-5° - 50°C (20° - 120°F) Measuring
-20° - 60°C (0° - 140°F) Ambient
Response Time:
T90 = 60 sec
Accuracy:
Max. Error < 2 % of measured range
Repeatability:
±0.5 % of measured range
Resolution:
0.01 ppm or 0.01 % Saturation
Operating Lifetime of Sensor Cap: Up to 2 years in recommended service, typically > 1 year
Wetted Materials:
316 SS, POM, PVC, Silicone
Sensor Cable:
Shielded 7 core cable, 7 meter (23 ft) or 15 meter (49 ft) lengths
Process Connection:
G1 Thread (¾” FNPT adapter available)
Maximum Cable Length:
100 m maximum from C-22 controller
Dimensions:
Length 8.7” (220 mm), Diameter 1.6” (40 mm)
64