beautypg.com

4 output signals, 5 contact relays, 6 display – Electro-Chemical Devices (ECD) T80 User Manual

Page 15: 7 enclosure, 8 power, 9 environmental conditions, 10 shipping

background image

Resistivity

2% of range

Turbidity

2% of reading

Temperature

± 0.3°C

1.2.4

O

UTPUT

S

IGNALS

4-20 mA output (standard, one per Channel), Fault Condition: 3.5 mA, 22 mA or none
Modbus RTU (standard)
HART® (optional)

1.2.5

C

ONTACT

R

ELAYS

(Optional) Three (3) SPDT, 1 form C, 250 VAC, 10 Amp resistive maximum, relays, user configurable as Hi/Lo
alarms with expiration timer, Periodic Timers or Fault alarms

1.2.6

D

ISPLAY

128 x 64 pixels (2.75” x 1.5”) LCD, Black on Grey background on loop powered instruments, Blue on White
background with LED backlight on 100-250 VAC and 24 VDC powered instruments, English or icon based
menus, numeric and graphical displays

1.2.7

E

NCLOSURE

Beige Polycarbonate, IP65, weatherproof, ½ DIN, (L x W x D) 5.7” X 5.7” X 3.5” (14.4cm X 14.4cm X 9.0cm)

1.2.8

P

OWER

Code -0

Loop powered, 24 VDC 600 Ω maximum load (18-36VDC @ 35 mW minimum)

Code -1

24 VDC (18-36 VDC @ 250 mW minimum)

Code -2

100-240 VAC, 50/60 Hz, 4W, protected with 250V, 1A, Slow Blow fuse

1.2.9

E

NVIRONMENTAL

C

ONDITIONS

Outdoor use (IP65)
Ambient Temperature

-20°C - 70°C (24 VDC Models)

-20°C - 60°C (100-240 VAC Models)

Storage Temperature

-30°C - 85°C

Relative Humidity

0 – 80%, up to 31°C

Decreasing linearly to 50% RH a 40°C

Altitude

Up to 2000 m (6500 ft)

Mains Supply Voltage

Fluctuations up to ±10% of the nominal voltage

Transient over voltages: CAT II

Pollution Degree: 2

1.2.10

S

HIPPING

Size

8” x 8” x 5” (20.5 cm x 20.5 cm x 12.7 cm)

Weight

1.6 lbs. (0.75 kg)

14

This manual is related to the following products: