Electro-Chemical Devices (ECD) T80 User Manual
Page 25

times and the State, energize: changes state from de-energized to
energized on alarm.
o
Timed, Enter Period, Duration times and Hold On/Off
o
Fault, No input required, relay condition changes from energize to de-
energize.
o
Disable, Inactivates relay and removes the relay 1 button from the
HOME Screen display.
HOLD, Freezes outputs at current value and locks relays in their current state.
•
Hold Timeout, Removes HOLD after a certain period
of time, default setting: No Timeout, selections
include 15 minutes, ½ hour, 1 hour
o
SERIAL MODBUS configure serial output,
ADDRESS, enter address: 001 to 247
BAUD, Choose baud rate, default 9600
FORMAT, set serial data format, default value: 8N1, 8 bit, no
parity bit, 1 stop bit
o
SERIAL HART configure output,
ADDRESS, enter address: 01-63
BAUD, default 1200, no adjustment available
FORMAT, default value: 8O1, 8 bit, Odd parity bit, 1 stop bit,
no adjustment available
o
PASSWD Enter 4 character password to protect access to MENU Level, CAL Menu, CONFIG
Menu and SIM Menu (simulate). Each level can be turned ON or OFF and can have a unique
password.
MENU
ON/OFF
_ _ _ _
Locks Main Menu
CAL
ON/OFF
_ _ _ _
Locks CAL and CONFIG
CONFIG
ON/OFF
_ _ _ _
Locks CONFIG
SIM
ON/OFF
_ _ _ _
Locks SIM and CONFIG
SENSOR enters the sensor configuration menu.
o
Choose SENSOR 1 or 2
TYPE, Allows T80 transmitter to configure the S80 sensor.
For use only when switching the measurement electrode
type in an S80 sensor, i.e. for a pH electrode to a pION
electrode. Select Sensor Type: pH, ORP, DO
2
, NH
3
, NH
4
+
, Br
-
,
Ca
++
, Cl
-
, Conductivity, Resistivity, Cu
++
, CN
-
, F
-
, NO
3
-
, K
+
, Ag
+
,
Na
+
, S
--
T COMP, Enter % temperature compensation per degree: pH, pION 0.33%, ORP 0.00%,
DO
2
4%, Conductivity 2%, Resistivity -5.4% (see Table Appendix D)
ISO PT, Enter Iso Potential value in mV. The Iso Potential is
the point where changes in the temperature do not cause
changes to the signal.
o
COMP Dual Channel Only, Sets compensation type: Dissociation
(pKa), NH
4
+
, Free Chlorine, HF, S
-2
, Interference, X ppm Sensor 1 = 1
ppm Sensor 2, Percentage % change per pH.
24