4.7 Instrument re-ranging

Re-ranging can easily be performed by changing the capacity of the instrument. This can be performed by

using FlowPlot, E-7000 or Bright readout/control module, RS232 digital communication or any available

fieldbus communication interface supporting parameter 21: capacity. Capacity can be set between minimum

capacity and maximum capacity. The actual values depend on the type of instrument.

FlowPlot can be downloaded at:

http://downloads.bronkhorst.com

Example of changing flow range using FlowPlot:

Enter max flow range here (capacity) or move the slider.

4.8 Instrument filter settings

Changing filter settings can be done using FlowPlot.

FlowPlot can be downloaded at:

http://downloads.bronkhorst.com

Output filter:

Dynamic display factor and Static display factor are filter settings for readout only (at analog output and digital

output). Output signal filtering at stable situation and at a flow-step can be set separately.

Sensor filter:

Exponential smoothing factor is the most important filter parameter to influence the stability of the meter and

will also influence the controller behaviour, where it will be provided as input for the controller.

Adaptive smoothing factor does not have functionality for a mini CORI-FLOW. For a mini CORI-FLOW it will

temporarily switch off the filter during a flow-step to enable fast controller behaviour together with strong

filtering.

For all filter settings:

A value can be entered between 0 and 1, where values close to 0 will filter strongly and slow down the output.

A value close to 1 will speed-up the output and will introduce more noise.

At value = 0, the filter has been switched off. No signal is put-through: no signal changing.

At value = 1, the output is equal to the input: no filtering at all.