4.3 Control system with XB-coil

For the XB-coil the calculation is as follows:

Because the capacity and inductivity are negligible, these can be left out of the calculation.

Resistance of the XB-coil = 300 Ω

Max. control voltage from the control circuit: 25 Vdc

25Vdc

300Ohm

XB

Control

circuit

Rk

Stahl

ISpac 9167

It has to be possible that the coil gets at least 11 Vdc, so:

I = 11

300

= 37 mA

The maximum output of the isolating repeater is 11,5 Vdc (at 25 Vdc input).

For cable losses is left : 11,5 - 11 = 0,5 Vdc

R

V

I

cable

=

=

=

0,5

0,037

13,5 Ω

0,25 mm

2

gives 1,6 Ω per 10 m, so the max. cable length is 13,5 x

1,6

10 = 84 m.

4.4 Standard cables for Bronkhorst High-Tech B.V. equipment

Cable length

10 m

20 m

60 m

100 m

Flow sensor

0,25 mm

2

0,25 mm

2

0,25 mm

2

0,25 mm

2

XB coil

0,25 mm

2

0,25 mm

2

0,25 mm

2

XC coil

0,25 mm

2

0,25 mm

2

NB: Cable capacity and inductivity are negligible.

4.5 EMC and cables

All instruments described in this manual carry the CE-mark.

Therefore they have to comply with the EMC requirements as are valid for this kind of instruments.

However compliance with the EMC requirements is not possible without the use of proper cables and

connector/packing gland assemblies.

For good results Bronkhorst High-Tech B.V. can provide standard cables. Otherwise follow the guidelines as

stated below.

Note: Readout units with its I.S. isolators should be situated outside zone 2.

4.5.1 Connector assembly customer side

Fold the shield of the cable back over the cable (the shield must be around the cable)

Wind a copper tape around the shield

Solder a black wire on the tape and

Connect to pin 9 of connector

copper tape

shielded cable

e.g. LAPP LiYCY

other wires

D-connector housing

metallized

connector

8 mm

20 mm

black wire

(shield)

page 20

9.17.028