MX Compliance Test System

System Administrator Manual

4.4.1

MXCTS Uncertainty - Harmonics

1

:

4 A RMS Range

Frequency

Uncertainty

RMS current 4

mA

Fundamental current

50 / 60 Hz

4 mA

Harmonic current

100 Hz- 1000 Hz

0.2 % Reading + 4 mA

1050 Hz - 2000 Hz

0.2 % Reading + 6 mA

2050 Hz - 2400 Hz

0.2 % Reading + 8 mA

16 A RMS Range

Frequency

Uncertainty

RMS current 16

mA

Fundamental current

50 / 60 Hz

16 mA

Harmonic current

100 Hz- 1000 Hz

0.2 % Reading + 4 mA

1050 Hz - 2000 Hz

0.2 % Reading + 6 mA

2050 Hz - 2400 Hz

0.2 % Reading + 8 mA

40 A RMS Range

Frequency

Uncertainty

RMS current 32

mA

Fundamental current

50 / 60 Hz

32 mA

Harmonic current

100 Hz- 1000 Hz

0.2 % Reading + 4 mA

1050 Hz - 2000 Hz

0.2 % Reading + 6 mA

2050 Hz - 2400 Hz

0.2 % Reading + 8 mA

75 A RMS Range

Frequency

Uncertainty

RMS current 64

mA

Fundamental current

50 / 60 Hz

64 mA

Harmonic current

100 Hz- 1000 Hz

0.2 % Reading + 4 mA

1050 Hz - 2000 Hz

0.2 % Reading + 6 mA

2050 Hz - 2400 Hz

0.2 % Reading + 8 mA

Table 4-1: MXCTS Measurement Uncertainty - Harmonics

1

Uncertainty data is given for products under test that exhibit a Power Factor of 0.75 < PF or PF < - 0.75.