MX Compliance Test System
System Administrator Manual
4.4.1
MXCTS Uncertainty - Harmonics
4 A RMS Range
Frequency
Uncertainty
RMS current 4
mA
Fundamental current
50 / 60 Hz
4 mA
Harmonic current
100 Hz- 1000 Hz
0.2 % Reading + 4 mA
1050 Hz - 2000 Hz
0.2 % Reading + 6 mA
2050 Hz - 2400 Hz
0.2 % Reading + 8 mA
16 A RMS Range
Frequency
Uncertainty
RMS current 16
mA
Fundamental current
50 / 60 Hz
16 mA
Harmonic current
100 Hz- 1000 Hz
0.2 % Reading + 4 mA
1050 Hz - 2000 Hz
0.2 % Reading + 6 mA
2050 Hz - 2400 Hz
0.2 % Reading + 8 mA
40 A RMS Range
Frequency
Uncertainty
RMS current 32
mA
Fundamental current
50 / 60 Hz
32 mA
Harmonic current
100 Hz- 1000 Hz
0.2 % Reading + 4 mA
1050 Hz - 2000 Hz
0.2 % Reading + 6 mA
2050 Hz - 2400 Hz
0.2 % Reading + 8 mA
75 A RMS Range
Frequency
Uncertainty
RMS current 64
mA
Fundamental current
50 / 60 Hz
64 mA
Harmonic current
100 Hz- 1000 Hz
0.2 % Reading + 4 mA
1050 Hz - 2000 Hz
0.2 % Reading + 6 mA
2050 Hz - 2400 Hz
0.2 % Reading + 8 mA
Table 4-1: MXCTS Measurement Uncertainty - Harmonics
1
Uncertainty data is given for products under test that exhibit a Power Factor of 0.75 < PF or PF < - 0.75.
Revision
B
25