beautypg.com

AMETEK MXCTSH Administrator Manual User Manual

Page 16

background image

MX Compliance Test System

System Administrator Manual

Field

Description

Harmonics calibration
coefficient.

Correction factor (in percent) used to compensate for frequency roll-off (if any)
at highest current harmonics.

Voltage Compensation

When on (set to 1), the CTS software will compensate for the effect of high
voltage distortion on the current harmonics measurements, if the VTHD is
found to be over the IEC limit, before applying the IEC test limits. This setting
should always be on (1).

Revision

B

16

This manual is related to the following products: