AMETEK MXCTSH Administrator Manual User Manual
Page 21

MX Compliance Test System
System Administrator Manual
3.4
Configuration Limits - Transitory State Harmonics - CTSMXH
This screen displays the Stage 2 Conditional Supply non balanced three phase (CSPNBTPE)
equipment permissible current harmonic limits. Note that each EUT Stage and type has a
corresponding limit screen.
The following connection types are supported using the abbreviations shown in the table:
Abbreviation
Meaning
CSPNBTPE
Connection Single Phase & Non Balanced Three Phase Equipment
CBTPE
Connection Balanced Three Phase Equipment
CBTPEUSC
Connection Balanced Three Phase Equipment Under Specified Conditions.
Table 3-1: Connection Type Abbreviations
If different Sce values apply to each category EUT, they will be shown between brackets
immediately following these abbreviations in the Limit drop down control.
The maximum number of harmonics supported is 50 but only 13 are presently covered by the IEC
61000-3-12 standard. Harmonics for which no limit applies are set to zero.
Revision
B
21