Error and warning messages
Error message
Cause
Solution
Error:Reading cal data.
Loading default data
Corruption on
calibration data. MPC
unit may be out of
calibration
Contact vendor, for receiving
instructions for re-calibrate the
instrument, if possible.
In
it
ia
li
za
ti
o
n
Error: Languaje file not
found!
Corruption on language
file
Reinstall the PDA software
ERROR: COM port error
An error has occurred
while connecting or
transferring data
Close the application and
restart the PDA and the MPC
unit
Error: Timeout opening port
Bluetooth COM port
failure
Close the application and
restart the PDA and the MPC
unit
C
o
m
m
u
n
ic
a
ti
o
n
s
Error: Port not opened
Bluetooth COM port is
closed
Close the application and
restart the PDA and the MPC
unit
Error: Invalid current value
Typed value is invalid
Type in a correct value
S
et
u
p
Error: Invalid file
File to open is invalid or
corrupted
Open a csv, MPC file
Error: Main contactor failure.
Keeps deenergized. Check
safety key
Safety key is in
“disabled” position
Turn the key to the “enabled”
position
Error: Cannot reach required
current. Continue
extrapolating data?
MPC unit is not able to
reach the requested
current, due to an
unidentified cause.
If allowed, PDA will
extrapolate the values to
the requested current,
but applying a lower
value.
Plug the MPC unit to a mains
socket connected to a mains
line with lower impedance, or
to a more powerful generator.
Try to reduce the load.
Danger: Main contactor
failure. Keeps energized
This is a life-
threatening scenario.
Main contactor is
broken, safety
protection is overriden
Contact vendor.
M
ea
su
ri
n
g
Erratic voltages higher than
50V. Continue?
This is a life-
threatening scenario.
Erratic voltages flowing
across the ground are
too high. A mains phase
may be tied to ground
Contact responsible of facility
under inspection.