Fault description

Possible fault cause

Required measures

Controller
No unit function

(dark display).

Display mode “Standby” active. Press any controller key.
Main power switch turned off. Turn on the main power switch.

No entries to controller keypad

possible. Notification “KEY

LOCK” is displayed

Keyboard locking (option) acti-

vated.

Unlock keyboard locking (chap.

14.3).

No access to menu “User set-

tings”“.

User code forgotten.

Contact BINDER service.

Wrong temperature alarms, dis-

turbance of temperature accuracy

Temperature unit changed to

°F.

Set temperature unit to °C (chap.

6.4).

Chart recorder function: meas-

ured-value memory cleared, in-

formation loss.

New setting of storage rate.

Change the storage rate ONLY if the

previously registered data are no

more required (chap. 7).

Controller does not attain set-

points entered in Manual Mode.

Button EXIT or AUTOMATIC

has been hit: Unit is in Idle

Mode.

Change to Manual Mode (chap. 8).

Controller does not attain pro-

gram set-points.

Button EXIT or AUTOMATIC

has been hit: Unit is in Idle

Mode.

Start the program again (chap. 9.7).

Program duration longer than

programmed.

Tolerances have been pro-

grammed.

For rapid transition phases, do NOT

program tolerance limits in order to

permit maximum heating, speed.

Program stops one section too

early.

Program line is incomplete.

When programming, define the end

value of the desired cycle by adding

an additional section with a section

time of at least one second.

RESET button does not cancel

the notifying or alarm indication.

Cause of disturbance not re-

moved correctly.

Remove cause of disturbance. If the

RESET button still does not cancel

the indication, contact BINDER ser-

vice.

Ramp temperature transitions are

only realized as steps.

When using the Program Edi-

tor of the software APT-

COM™ 3 DataControlSystem,

the setting “step” has been

selected.

Select setting „ramp“ in the Program

Editor of the software APT-COM™ 3

DataControlSystem and transfer a

program to the chamber controller.

Display flashing:

1999 or -1999 or 9999.

Sensor rupture between sen-

sor and controller or Pt 100

sensor defective.

Contact BINDER service.

Short-circuit.
Initialization problem due to

turning on the chamber too

early.

Observe a delay time of about 30s

between turning Off and On again

the chamber.

Miscellaneous
Fan does not turn

Fan speed set to 0%..

Set the fan speed to the desired

value.

Repair must only be performed by qualified service personnel authorized by BINDER.

Repaired units must comply with the BINDER quality standards.