3 lateral control panel mk – BINDER MK 53 User Manual
Page 17

MK (E2.1) 10/2014
page 17/73
2.3 Lateral control panel MK
Figure 5: Lateral control panel at the left side of the refrigerating machine
with options over-/under temperature safety device, and analog output
(3) Free place for optional button
(4) Switch for interior chamber light on / off
(5) RESET button for option over-/under temperature safety device class 2 (option, chap. 11.3)
(6) Over-/under temperature safety device (option, chap. 11.3)
(6a) Module for upper temperature limit
(6b) Module for lower temperature limit
(7) RS 422 serial interface for computer communication
(8) Analog output (option)
The current temperature value is put out as analogue signal 4-20 mA to a DIN socket. Measuring
range 4 mA = -40 °C, 20 mA = +180 °C.
(9) Not used
RESET CL. 2.0
(5)
(6a)
(6b)
(7)
(8)
(9)
(4)
(3)