BINDER M 53 User Manual
Page 58
![background image](https://www.manualsdir.com/files/767105/content/doc058.png)
M (E2) 10/2014
page 58/71
Unit size
53
115
240
400
720
Temperature data (continued)
Recovery time after door
was opened for 30 sec 2)
to 70 °C
min
1
1
1
1
1
to 150 °C
min
3
3
3
3
3
to 300 °C
min
5
5
5
5
5
Air change (air flap open)
at 70 °C
x/h
180
87
57
51
33
at 150 °C
x/h
192
96
60
54
36
at 300 °C
x/h
160
78
54
48
29
Electrical data
IP system of protection acc. to EN 60529
20
20
20
20
20
Nominal voltage (±10 %) 50/60 Hz
V
230 1N~ 230 1N~ 230 1N~ 400 3N~ 400 3N~
Nominal power
kW
1.20
1.60
2.70
3.40
5.00
Energy consumption 3)
at 70 °C
Wh/h
145
230
370
520
570
at 150 °C
Wh/h
300
544
850
1200
1320
at 300 °C
Wh/h
720
1100
1400
2340
2600
Power plug
shock
proof plug
shock
proof plug
shock
proof plug
CEE plug
5 poles
CEE plug
5 poles
Installation category acc. to IEC 61010-1
II
II
II
II
II
Pollution degree acc. to IEC 61010-1
2
2
2
2
2
Legend:
1) Door without window
2) Up to 98 % of the set value
3) Use this value for sizing air condition systems.
All technical data is specified for unloaded units with standard equipment at an ambient temperature of
+25 °C / 77 °F and a power supply voltage fluctuation of ±10. The temperature data is determined in ac-
cordance to BINDER factory standard following DIN 12880, observing the recommended wall clearances
of 10 % of the height, width and depth of the inner chamber. Technical data refers to 100% fan speed.
All indications are average values, typical for units produced in series. We reserve the right to
change technical specifications at any time.
If the chamber is fully loaded, the specified heating up and cooling down times may vary ac-
cording to the load.