DANGER
Electrical hazard.
Danger of death.
The unit must NOT become wet during operation or maintenance.
The drying and heating ovens with forced convection were produced in accordance with VDE regulations
and were routinely tested in accordance to VDE 0411-1 (IEC 61010-1).
During and shortly after operation, the temperature of the inner surfaces almost equals the set-point.
CAUTION
The inner chamber, the door gasket, and the exhaust duct will become hot during
operation.
Danger of burning.
Do NOT touch the inner surfaces, the door gasket, the exhaust duct or the charging
material during operation.
1.6 Intended use
Drying and heating ovens with forced convection and advanced program functions M are suitable for
drying and heat treatment of solid or pulverized charging material, as well as bulk material, using the
supply of heat. They are suitable for harmless materials. A mixture of any component of the charging
material with air must NOT be explosive. The operating temperature must lie below the flash point or
below the sublimation point of the charging material.
Other applications are not approved.
Drying and heating ovens with forced convection and advanced program functions M are not
classified as medical devices as defined by the Medical Device Directive 93/42/EEC.
Drying and heating ovens with forced convection M can be used for drying purposes but are specially
designed for solving all the problems which occur during material and ageing tests.
Do NOT use the unit for drying processes when large quantities of vapor would form and result in con-
densation.
Due to the special demands of the Medical Device Directive (MDD), these ovens are not
qualified for sterilization of medical devices as defined by the directive 93/42/EWG.
Following the instructions in this operating manual and conducting regular maintenance work
(chap. 12) are part of the intended use.
The charging material shall not contain any corrosive ingredients that may damage the ma-
chine components made of stainless steel, aluminum, and copper. Such ingredients include
in particular acids and halides. Any corrosive damage caused by such ingredients is exclud-
ed from liability by BINDER GmbH.