Start up, 2 operating modes – BINDER M 53 User Manual
Page 19
M (E2) 10/2014
page 19/71
5.
Start up
After connecting the electrical supply (chap. 4.1), turn on the unit via the main power switch (2).
Warming chambers may release odors in the first few days after commissioning. This is not a quality de-
fect. To reduce odors quickly we recommend heating up the chamber to its nominal temperature for one
day and in a well-ventilated location.
5.1 Function overview of the MB1 display program controller
Pilot lamp: Ready for operation
EXIT button (to exit a menu point)
AUTOMATIC button (to start a previously entered program)
ENTER button (to confirm a selection)
Navigation buttons (functions are assigned by the menu)
Figure 5: Display program controller MB1
The program controller MB1 controls the temperature inside the drying and heating oven (range: 5 °C
above ambient temperature up to 300 °C).
You can enter the desired set point values in Manual Mode or Program Mode (chap. 5.2) in the display
controller.
W
X
08:43:55 15.12.13
CONFIG
VIEW->
PGM
TEMP
40.0
36.8
°C
HAND
Set point value Actual value
Figure 6: Normal display of the MB1 program controller in Manual mode
5.2 Operating modes
The program controller MB1 operates in 3 modes:
Idle Mode
The controller is not functional, i.e., there is no heating. The fan turns
at a 50% rate.
Manual Mode
(Fixed value operation) (HAND)
The controller operates as a fixed-point control, i.e., a temperature
set-point can be defined, which is then maintained (chap. 8).
Program Mode (AUTO)
An entered temperature program is run (chap. 9).
The program controller MB1 allows programming temperature cycles.
The controller offers 25 program memory positions with 100 program sections each. The total number of
program sections of all programs is limited to 500.