page 36/55

CAUTION

Use of inappropriate plug.
Danger of inflammation.

To close the exhaust duct use the delivered plug only.

For drying purpose, please remove the plug in order to permit dissipation of the generated vapor, which

would lead to condensation in the inner chamber.

8.6 Analog output for temperature (option)

With this option the chamber is equipped with an analog output 4-20 mA for temperature. This output

permits transmitting data to external data registration systems or devices.
The connection is carried out as a DIN socket at the rear of the chamber as follows:

ANALOG OUTPUT 4-20 mA DC

PIN 1: temperature –
PIN 2: temperature +

Temperature range:
0 °C to +300 °C

A suitable DIN plug is enclosed.

Figure 9: Pin allocation of DIN socket for option analogue outputs

9.

Maintenance, cleaning, and service

9.1 Maintenance intervals, service

DANGER

Electrical hazard.
Danger of death.

The unit must NOT become wet during operation or maintenance work.

Do NOT remove the rear panel of the unit.

Before conducting maintenance work, turn off the unit at the main power switch and

disconnect the power plug.

Ensure all maintenance work is conducted by licensed electricians or experts authorized

by BINDER.

Ensure regular maintenance work is performed at least once a year.

The warranty becomes void if maintenance work is conducted by non-authorized personnel.

Replace the door gasket only when cold. Otherwise, the door gasket may become damaged.