FED (E2) 02/2015

page 33/55

8.

Options

8.1 Communication software APT-COM™ 3 DataControlSystem (option)

The oven is regularly equipped with a serial interface RS 422 that can connect the BINDER

communication software APT-COM™ 3 DataControlSystem. The connection to a computer is established

using the FED interface via an interface converter RS 422 / RS 232.

Make sure that the interface mode is correctly set to

“0” = “Modbus” in the user level (chap.

6.4.5).

The actual temperature, and fan speed values are given at adjustable intervals. Programming can be

performed graphically via PC. Up to 30 chambers with RS 422 interface can be cross linked. For further

information, refer to the operating manual of the BINDER communication software APT-COM™.
Pin allocation of the RS 422 interface: pin 2: RxD (+)

pin 3: TxD (+)
pin 4: RxD (-)
pin 5: TxD (-)
pin 7: Ground

If several drying and heating ovens FED are to be recorded via a PC, each one must be

allocated a unique address. Addressing is performed via the chamber controller (see chap.

6.4.3).


8.2 Data logger kit

BINDER Data Logger Kits offer an independent long-term measuring system for temperature. They are

equipped with a keyboard and a large LCD display, alarm functions and a real-time function.

Measurement data are recorded in the Data Logger and can be read out after the measurement via the

RS232 interface of the Data Logger. It offers a programmable measuring interval and permits storing up to

64000 measuring values. Reading out is done with the Data Logger evaluation software. You can give out

a combined alarm and status protocol directly to a serial printer.
Data Logger Kit T 350: Temperature range 0 °C / 32 °F up to +350 °C / 662 °F

For detailed information on installation and operation of the BINDER Data Logger, please

refer to the mounting instructions Art. No. 7001-0204 and to the original user manual of the

manufacturer, supplied with the data logger.


8.3 HEPA fresh air filter (option)

With this option, the introduced fresh air is cleaned by means of a high efficiency submicron particulate air

filter type HEPA class H 14 (acc. to DIN EN 1822). Replace the filter insert, if necessary, by removing the

metal cover of the filter at the left side of the unit (Art. No. 6014-0003).