General information for safe handling of gas cylinders:

Store and use gas cylinders only in well ventilated areas.

Open the gas cylinder valve slowly to avoid pressure surges

Secure gas cylinders during storage and use against falling (chaining).

Transport gas cylinders with a cylinder cart, do not carry, roll, or throw them

Always close the valve even with apparently empty cylinders; screw on the cap when not in

use. Return gas cylinders with the valve closed

Do not open gas cylinders by force. Mark them when damaged

Observe relevant regulations for dealing with gas cylinders.

Connect a flexible gas tube to the gas hose connection adapter (diameter 10mm), which is used for gas

inlet, and secure it with hose clamps (hose and hose clamps are not enclosed). There is a constant gas

flow after establishing the connection.

After connecting the gas cylinder, check all gas connections for leaks (e.g. with leak spray or

diluted soap solution).

Use a pressure reducer and make sure to avoid any excessive outlet pressure when connecting the gas

hose to the oven.

The unit is not entirely gas-tight. Inert gases from inside the drying and heating oven can

escape into the surrounding atmosphere.

Inert gases in high concentrations is hazardous to health. They are colorless and almost odorless and

therefore practically imperceptible. Inhalation of inert gases can cause drowsiness up to respiratory arrest.

When the O

2

content of the air decreases below 18%, there is risk of death from lack of oxygen. Any gas

that might escape has to be led out via good room ventilation or a suitable exhaust system.

WARNING

High concentration of inert gas.
Risk of death by suffocation.

Do NOT set up units in non-ventilated recesses.

Ensure technical ventilation measures.

Respect the relevant regulations for handling these gases.

Inert gases, which are heavier than air, may accumulate in low-lying areas of the installation

site.

The “Mostly gas-tight version” reduces the loss of gas.
Setting (sample values):

If you want to flush the unit with an air exchange rate of 1 per hour, set the flow rate on the pressure

reducer according to the interior volume.
FED 53 with 53 l internal volume: The flow rate corresponding to 53 l / h is 0.9 l / min.
FED 115 with 115 l internal volume: The flow rate corresponding to 115 l / h is 1.9 l / min.

The air flap does not close the exhaust duct completely. The delivered plug serves to avoid loss of

introduced inert gas via the exhaust duct. Due to special demands of heat resistance, use the delivered

plug only.