BINDER BD 23 User Manual
Page 51

BD / ED / FD (E2) 02/2015
page 51/63
Unit size
FD 23
FD 53
FD 115
FD 240
Temperature data
Temperature range, 5 °C above ambient
up to
°C / °F
300 / 572 300 / 572 300 / 572 300 / 572
Temperature fluctuation at 150 °C / 302 °F
± K
0.3
0.3
0.3
0.3
Temperature
uniformity
(variation 1)
at 70 °C / 158 °F
± K
0.8
0.8
0.7
0.8
at 150 °C / 302 °F
± K
2.2
2
1.8
2
at 300 °C / 572 °F
± K
4.3
3.7
3.9
4.3
Heating up time 2)
to 70 °C / 158 °F
Min
7
7
7
11
to 150 °C / 302 °F
Min
22
24
28
24
to 250 °C / 482 °F
Min
45
60
49
50
Recovery time
after door was
open for 30 sec
2)
at 70 °C / 158 °F
Min
2
2
2
2
at 150 °C / 302 °F
Min
4
5
5
6
at 300 °C / 572 °F
Min
9
9
12
13
Ventilation data
Air change
at 70 °C / 158 °F
x/h
59
59
29
19
at 150 °C / 302 °F
x/h
64
64
32
20
at 300 °C / 572 °F
x/h
53
53
26
18
Electrical data
IP system of protection acc. to EN 60529
IP
20
20
20
20
Nominal voltage (±10%) 50/60 Hz
V
230 1N~ 230 1N~ 230 1N~ 230 1N~
Nominal power
kW
0.8
1.20
1.60
2.70
Power plug
shock proof plug
Unit fuse
5x20mm / 230V / time-lag M
Amp
10
external
10
external
10
external
16
external
Energy
consumption
at 70 °C / 158 °F
Wh/h
145
172
230
370
at 150 °C / 302 °F
Wh/h
300
429
544
850
at 300 °C / 572 °F
Wh/h
720
951
1100
1400
Installation category acc. to IEC 61010-1
II
II
II
II
Pollution degree acc. to IEC 61010-1
2
2
2
2
Electrical connection data FD-UL acc. to cUL standard (for USA and Canada):
Unit size
FD 23-UL FD 53-UL FD 115-UL FD 240-UL
Electrical data
Nominal voltage (±10%) 60 Hz
V
115 1N~ 115 1N~ 115 1N~ 208 3N~
Nominal power
kW
0.80
1.20
1.60
2.70
Nominal current
Amp
7.0
10.5
14.0
11.2
Power plug
NEMA
5-15P
5-20P
5-20P
L21-20P
Unit fuse
6,3 x 32 mm / 250V / super time-lag TT
Amp
12,5
external
16
external
16
external
3 x 16
internal
Legend: 1) without outer glass door 2) up to 98 % of the set value
All technical data is specified for unloaded units with standard equipment at an ambient temperature of
+25 °C / 77 °F and a power supply voltage fluctuation of ±10. The temperature data is determined in
accordance to BINDER factory standard following DIN 12880, observing the recommended wall
clearances of 10 % of the height, width and depth of the inner chamber. Technical data refer to 100% fan
speed.
All indications are average values, typical for units produced in series. We reserve the right to
change technical specifications at any time.
If the cabinet is fully loaded, the specified heating up times may vary according to the load.