Danger, Caution – BINDER BD 23 User Manual
Page 16
BD / ED / FD (E2) 02/2015
page 16/63
DANGER
Explosion hazard.
Danger of death.
∅
Do NOT introduce any substance into the incubator / drying and heating oven which is
combustible or explosive at working temperature.
∅
NO explosive dust or air-solvent mixture in the inner chamber.
Any solvent contained in the charging material must not be explosive or inflammable. I.e., irrespective of
the solvent concentration in the steam room, NO explosive mixture with air must form. The temperature
inside the chamber must lie below the flash point or below the sublimation point of the charging material.
Familiarize yourself with the physical and chemical properties of the charging material, as well as the
contained moisture constituent and its behavior with the addition of heat energy.
Familiarize yourself with any potential health risks caused by the charging material, the contained
moisture constituent or by reaction products that may arise during the temperature process. Take
adequate measures to exclude such risks prior to putting the unit into operation.
DANGER
Electrical hazard.
Danger of death.
∅
The unit must NOT become wet during operation or maintenance.
The incubators BD and drying and heating ovens ED and FD were produced in accordance with VDE
regulations and were routinely tested in accordance to VDE 0411-1 (IEC 61010-1).
During and shortly after operation, the temperature of the inner surfaces almost equals the set-point.
CAUTION
The glass doors and glass door handles (BD), inner chamber, exhaust duct, door
window (option), and the door gaskets will become hot during operation.
Danger of burning.
∅
Do NOT touch the glass doors, inner surfaces, exhaust duct, door window, access
ports, door gaskets, or the charging material during operation.