Installation of the equipment, 1 electrical connection, Caution – BINDER BD 23 User Manual
Page 22

BD / ED / FD (E2) 02/2015
page 22/63
4.
Installation of the equipment
4.1 Electrical connection
The incubators BD and drying and heating ovens ED / FD are supplied ready for connection
•
BD all volumes, ED up to volume 240, FD all volumes:
Shockproof plug, power supply voltage 230 V (1N~) +/- 10 %, 50/60 Hz
Fixed power connection cable of 1800 mm in length
•
ED 400, ED 720:
CEE plug 5 poles, power supply voltage 400 V (3N~) +/- 10 %, 50/60 Hz
Fixed power connection cable of 2700 mm in length
•
BD-UL 23, BD-UL 53, BD-UL 115, BD-UL 240, BD-UL 400, ED 23-UL, FD 23-UL:
NEMA plug 5-15P, power supply voltage 115 V (1N~) +/- 10 %, 60 Hz
Fixed power connection cable of 1800 mm in length
•
BD 720-UL, ED 53-UL, ED 115-UL, FD 53-UL, FD 115-UL:
NEMA plug 5-20P, power supply voltage 115 V (1N~) +/- 10 %, 60 Hz
Fixed power connection cable of 1800 mm in length
•
ED 240-UL, ED 400-UL, ED 720-UL, FD 240-UL:
NEMA plug L21-20P, power supply voltage 208 V (3N~) +/- 10 %, 60 Hz
Fixed power connection cable of 2700 mm in length
•
The socket must also provide a protective conductor.
•
Prior to connection and start-up, check the power supply voltage. Compare the values to the specified
data located on the unit’s type plate (unit front behind the door bottom left-hand, or on the left-hand
side of the unit, chap. 1.4). We recommend the use of a residual current circuit breaker.
•
When connecting, please observe the regulations specified by the local electricity supply company
and as well as the VDE directives (for Germany)
•
Pollution degree (acc. to IEC 61010-1): 2
•
Over-voltage category (acc. to IEC 61010-1): II
CAUTION
Danger of incorrect power supply voltage.
Damage to the equipment.
Check the power supply voltage before connection and start-up.
Compare the power supply voltage with the data indicated on the type plate.
See also electrical data (chap. 12.4 to 12.6).
To completely separate the unit from the power supply, you must disconnect the power plug.
Install the unit in a way that the power plug is easily accessible and can be easily pulled in
case of danger.