beautypg.com

SOR 534CR Pressure Transmitter User Manual

Page 3

background image

Form 903 (03.13) ©SOR Inc.

3/8

Power

Supply

10 - 30 VDC

milliammeter

4 - 20 mA

Red

Black

Case

Ground

Zero

Not Used

Span

Bare

Connect the transmitter as shown

in . Case ground must be

connected to earth ground to

ensure EMI/RFI protection.

Apply pressure at which 4 mA output

is desired. (Zero may be adjusted
±10% of the upper range limit.)

With pressure source steady at the

desired zero level, rotate the zero
adjustment (#1) for a 4 mA indication
on the milliammeter.

Apply pressure at which 20 mA out-

put is desired. Span may be adjusted

from 25 to 100% of the upper range limit. (Maximum turndown is 4:1.)

With pressure source steady at the desired span level, rotate the span adjustment (#2)

for a 20 mA indication on the milliammeter.

Repeat steps 2 through 5 as needed if offsetting 4 mA from the normal zero point.

If interaction occurs, turn zero and span 15 turns counterclockwise. Repeat steps 2
through 6 above.

Calibration Procedure
The zero and span calibration procedure should be performed under ambient process
temperature conditions.

A pressure source with a calibrated reference gage, a milliammeter and a DC voltage
supply are required. Note the adjustable range on the instrument nameplate. For both
zero and span adjustments, turn the adjustment screw clockwise to increase,
counterclockwise to decrease.