4555 - 58 Series
5G & 10L Pressure Reactors
- 24 -
10. To operate reactors equipped with an air motor, connect air hose to a compressed air
line. For best torque and speed control the piping to the motor should be at least 3/8" IPS or
larger. Fill the oiler with SAE 10 oil and adjust the oiler feed to a rate of one drop per minute
into the air stream. For long continuous runs at high speeds, the oiling rate should be
increased to three drops per minute.
If the motor becomes sluggish, flush it with a non-flammable solvent in a well ventilated
area. Disconnect the air line and muffler and pour a small amount of solvent into the inlet
port. Rotate the shaft by hand in both directions for a few minutes; then connect the air line
and run the motor until there is not further trace of solvent in the exhaust. If the muffler is
dirty, replace it. Re-lubricate the motor with a squirt of oil into the chamber and reassemble.
11. If servicing assistance is needed, contact Parr Instrument Company directly at the
address shown on the back of these instructions
Fuses
The replacement of protective fuses should be performed by qualified personnel.
Note: Check the labels on the instrument for correct fuse rating.
Parr Part #
Description
Slave Box
Type
Ratings
1535E
Heater Fuse
A1761HC3
KAC
25 Amps, 600VAC
185E
Master
Controller Fuse
A1761HC3
Slo-Blo
5 Amps, 250VAC
2094E
Heater Fuse
A1761HCEE Fast-Acting 25 Amps, 250VAC
2093E
Main Fuse
A1761HCEE Fast-Acting 50 Amps. 250 VAC
139E8 Lift
Fuse A1761HCEE
Slo-Blo 2.5
Amps,
250VAC