8 Maintenance and service

Version 04/2012, Rev. 1.3 of 25/11/2013 • sb

51

Translation of the original operating manual

Pos: 159 /20 0 Ch rist/3 71 RVC-BA (PROJEK TE)/RVC 2- 18 CD plus_ 2-1 8 CDplus _HCl/08 0 Wa rtun g un d In stan dhalt ung/ 080 -00 10- 002 0 Ro tork am mer @ 9\ mo d_13 206 573 892 32_ 68.d ocx @ 540 11 @ 3 @ 1

8.1.2 Rotor chamber

If the rotor chamber is contaminated or if glass is broken, the rotor
chamber, the rotor shaft and the rotor must be cleaned immediately in order
to prevent corrosion and damage to the ball bearings.

Cleaning of the rotor chamber


1

Rotor chamber

2

Cap

3

Rotor shaft

Fig. 41: Rotor chamber with rotor shaft


• Open the lid.
• Disconnect the power cord from the wall outlet or instrument receptacle

before cleaning.

DANGER

Hot surface!
Let the rotor chamber cool down to room temperature!


• Pull off the cap.
• Remove the countersunk screw with the supplied hexagon socket key

(size 2.5).