MAINTENANCE

Replacing Blood System Filter

If reduced blood flow is experienced, the filter may be clogged and
needs replacement.

Never run the Patient Simulator without a filter.

Bloo

d R

igh

t L

eg

1 Turn off the Patient Simulator.

2 Remove genitalia with the catheterization assembly for easy

access.

3 Disconnect the filter from the right leg and pelvis blood tubes

and remove it.

4 Connect a new filter by reversing the steps above.

Servicing the Patient Simulator

A full service, including cleaning of the base board, should be
performed at regular intervals.

Caution: All servicing must be performed by qualified service

personnel.

Always Perform a Service:

− If liquids have been spilled in the Patient Simulator

− After use in dusty environments.

Caution: Do not use cables or connectors showing visible damage.