Chapter 6: Using The RS232 Receptacle

Product Service 1-800-522-7658

31

The time between data transmissions may be varied by the user to occur at 10, 30,
60, 300 or 600 second intervals. Press the MENU button on the Freeze Dryer
base unit until the RS232 screen appears on the display.





Press select until the desired time interval is shown.

The format of the transmitted message with typical values is as follows:



Where:

B

= Base Unit

T

= Collector Temperature

-48

= Collector Temperature in °Celsius

V

= Vacuum

18

=Vacuum Level in Micro Bars

SF

= Shell Freezer Temperature (if so equipped)

-41

= Shell Freezer Temperature in °C

C

=Vacuum Chamber Temperature (if so equipped)

9

=Feature not installed or out of range

PG

= Purge (if so equipped)

1

= ON

0

= OFF

TD

= Tray Dryer

PR

= Program

1

= Program 1 is selected

S

= Segment

3

= Segment 3 is operating

R

= Segment is in Ramp Mode

H

= Segment is in Hold Mode

TR

= Time Remaining of the current Hold Segment

5678

= Time Remaining in Hours

ST

= System Temperature

-40

= System Temperature in °C

P1

= Probe 1 Temperature

-39

= Probe 1 Temperature in °C

P2

= Probe 2 Temperature

999

= Probe not connected or temperature is out of range

P3

= Probe 3 Temperature

-37

= Probe 3 Temperature in °C

V

= Vacuum at Stoppering Tray Dryer

537

= Vacuum in Micro Bars

RS-232 TRANSMISSION RATE

10

SECONDS

B TD