Chapter 6: Maintaining Your Filtered Enclosure

Product Service 1-800-522-7658

50

Initial Certification

The filtered enclosure has been certified at the factory for an inflow velocity of
90±10 fpm along with the HEPA Filter Leak Test. The filtered enclosure should
be certified for the proper inflow velocity required by your Safety Officer. It is
also a conservative recommendation to perform the HEPA Filter Leak Test again
should there be any damage caused during transport.

Re-Certification

Under normal operating conditions, the enclosure should be recertified at least
annually if serviced. The certifier should perform the following tests.

Inflow Velocity Test

HEPA Filter Leak Test


In addition, the following tests should also be performed at the user’s discretion:

Measure of Line Voltage and Current

Smoke Test to determine proper airflow patterns

Lighting Intensity Test (when appropriate)

Noise Level Test (when appropriate)

Vibration Test (when appropriate)

Fluorescent Light Replacement

1. Disconnect the power.

2. Locate the small light reflector located under the control panel shown in

Figure 6-2 and Figure 6-3.

3. Remove the light reflector support by removing two Phillips screws on the

bottom of the light reflector.

4. Rotate and remove the old fluorescent lamp.

5. Reinstall the new fluorescent lamp and light reflector in reverse order.

6. Power the unit up and try the new fluorescent lamp.