Chapter 5:

Using Your Filtered Enclosure

Now that the installation of your filtered enclosure is completed, you are ready to
use your filtered enclosure.

Read this chapter to learn about:

Routine Daily Work Procedures.

Suitable Applications.

Appropriate HEPA Filter Applications, Suitability and Guidelines.

Odor Control Carbon Filter Applications.

Definition of Terms.

Appropriate Chemicals for Odor Control Carbon Filters.

Hazardous Misapplications for Odor Control Carbon Filters.

Chemical Carcinogen use with Odor Control Carbon Filters.

Prohibited Acid Use.

Routine Daily Work Procedures

Planning

Thoroughly understand procedures and equipment required before

beginning work. Balances typically work best with the draft shield closed
tightly and the enclosure set between 60-90 fpm.

Arrange for minimal disruptions, such as room traffic or entry into the

room while the enclosure is in use.

Start-up

Turn on exhaust system and accessory light.

Only raise the sash for loading and cleaning.

Check the baffle air slots for obstructions.

Allow the enclosure to operate unobstructed for 1 minute.

Wear a long sleeved lab coat and rubber gloves. Use protective eyewear.

Wear a protective mask if appropriate. Consult your Safety Officer for
additional personal protective equipment recommendations.