Chapter 2: Prerequisites

Product Service 1-800-522-7658

6

Do not position the unit so that it is difficult to operate the main
disconnect device.

Ne placez pas l'appareil de sorte qu'il est difficile de faire
fonctionner le dispositif principal de déconnexion.

Exhaust and Blower Requirements

The XPert Filtered Balance System uses an integral motorized impeller to draw
room air past the operator and through the enclosure. This contaminated air is then
pushed through the HEPA filter. The HEPA-filtered exhaust air is then forced out
the top of the enclosure. An optional carbon filter may be installed on the
downstream side of the HEPA filter to protect against nuisance odors.

The HEPA-filtered exhaust air can be recirculated into the laboratory or exhausted
outside with the addition of the Canopy Exhaust Connection and remote blower
listed in Chapter 7. Electrical connections are covered in Chapter 2 for this
configuration.

The XPert Filtered Balance Station does not include the motorized impeller and
must be exhausted to the outside by a remote blower. The XPert Filtered Balance
Station remote blower may be switched from the blower switch on the enclosure.
See Electrical Requirements in Chapter 2, and Electrical Supply Connections in
Chapter 3. Only one 6" exhaust connection is required for the 2', 3', or 4' XPert
Filtered Balance Stations to exhaust to the outside. Only one 10" exhaust
connection is required for the 5' and 6' XPert Filtered Balance Stations.

Data for the exhaust volume, noise pressure and enclosure static pressure loss are
listed for each filtered enclosure model at face velocities of 60, 80, and 100 fpm.
Balances typically work best between 60-90 fpm with the balance shield closed
tightly.

Enclosure

Width and

Sash

Opening

Model Description

Face Velocity

(fpm)

Exhaust

Volume

(CFM)

XPert Station

Initial Static

Pressure Loss

with HEPA

filter (in w.g.)

XPert

System

Noise

Pressure db

(A)

Xpert System

Max.

Equivalent

Resistance (Ft)

Xpert System

Max. External

Static Pressure

2'

9.4" Sash

2’ Xpert Filtered Balance System

2’ Xpert Filtered Balance Station

(6" Duct)

60
80

100

85

115
145

.35"- .41"

.48" - .57"
.62" - .72"

48-53
50-56
53-58

400
200

80

.20"
.18"
.13"

3'

9.4" Sash

3’ Xpert Filtered Balance System

3’ Xpert Filtered Balance Station

(6" Duct)

60
80

100

130
175
220

.37" - .43"
.53" - .62"
.68" - .79"

48-53
53-57
58-61

130

50
25

.15"
.11"
.08"

4'

9.4" Sash

4' XPert Filtered Balance System

4' XPert Filtered Balance Station

(6" Duct)

60
80

100

175
235
295

.40" - .46"
.57" - .66"

.76" - .87".

49-54
58-61
62-66

100

50
25

.20"
.18"
.13"

4'

12" Sash

4' XPert Filtered Balance System

4' XPert Filtered Balance Station

(6" Duct, Tall or Deep Models)

60
80

100

230
305
380

.56" - .65"
.77" - .87"

.92" - 1.04".

58-61
62-66
67-70

50
25

5

.18"
.13"
.05"

5'

12" Sash

5' XPert Filtered Balance System

5' XPert Filtered Balance Station

(10" Duct)

60
80

100

290
385
480

.38" - .46"
.55" - .67"
.72" - .85"

52-56
58-63
63-67

450
230
120

.22"
.20"
.15"

6'

12" Sash

6' XPert Filtered Balance System

6' XPert Filtered Balance Station

(10" Duct)

60
80

100

350
465
580

.46" - .56"
.67" - .81"

.87" – 1.03"

55-57
61-63
67-69

300
150

70

.20"
.18"
.13"

!