CE DECLARATION OF CONFORMITY

Recommendations for Users

Before using these products, make sure that they are entirely suitable for the environ-
ment in which they are used.

Operation of these instruments in residential areas could cause unacceptable interfer-
ence to radio and TV equipment.

Any variation introduced by the user to the supplied equipment may degrade the
instruments' EMC performance.

Unplug the instruments from power supply before replacing the fuse or making any
electrical connections.

CALIBRATION TEST KEYS

HANNA test keys provide a quick and easy way to test the meter's
accuracy by simply plug the key to the probe socket on the meter.
If the reading differs by more than ±0.4°C (±0.8°F) from the test
key value, the unit is due for recalibration.
Choose the proper test key according to your application needs.

HI 762-18C

Calibration test key,at -18.0°C

HI 762000C

Calibration test key, at 0.0°C

HI 762070C

Calibration test key, at 70.0°C

HI 762-004F Calibration test key, at -0.4°F

HI 762032F

Calibration test key, at 32.0°F

HI 762158F

Calibration test key, at 158.0°F

HI 7101

RH calibration chamber

HI 7102

RH calibration chamber for probes with sintered cap

HI 7111/P

Spare saturation LiCl salts for low humidity calibra-
tion (15 g, 6 pcs)

HI 7121/P

Spare saturation NaCl salts for high humidity cali-
bration (33 g, 6 pcs)