• Add the content of one packet of HI 93751-0

Indicator reagent.

• Replace the cap and invert gently to mix for

1 minute (about 30 inversions).

• Reinsert the cuvette into the instrument.

• Press Timer and the display will show the countdown prior to the measurement or, alternatively, wait

for 5 minutes and press Read. When the timer ends the meter will perform the reading.

• The instrument displays concentration in mg/L of Sulfate (SO

4

2

¯).

INTERFERENCES

Interferences may be caused by:
Calcium (as CaCO

3

) above 20000 mg/L

Chloride (as Cl

-

) above 40000 mg/L

Magnesium (as MgCO

3

) above 10000 mg/L

Silica (as SiO

2

) above 500 mg/L

Color or suspended matter in large amounts will interfere: suspended matter should be removed by previous
filtration.
Organic matter in large amounts may impede the precipitation of barium sulfate.

Sulfate