Hanna Instruments HI 83225 User Manual
Page 40
40
Phosphorus LR
PHOSPHORUS LOW RANGE
SPECIFICATIONS
Range
0.0 to 10.0 mg/L
Resolution
0.1 mg/L
Accuracy
±0.5 mg/L ±4% of reading at 25 °C
Typical EMC Dev. ±0.1 mg/L
Light Source
Tungsten lamp with narrow band interference filter @ 525 nm
Method
Adaptation of the
Standard Methods for the Examination of Water and Wastewater,
18
th
edition, Amino Acid method. The reaction between phosphate and reagents
causes a blue tint in the sample.
REQUIRED REAGENTS
Code
Description
Quantity
HI 93706A-0
Molybdate
10 drops
HI 93706B-0
Amino Acid Powder
1 packet
REAGENT SETS
HI 93706-01 Reagents for 100 tests
HI 93706-03 Reagents for 300 tests
For other accessories see page 57.
MEASUREMENT PROCEDURE
Note: for sample preparation follow the IRRIGATION WATER (LR) procedure at page 20.
• Select the
Phosphorus method using the procedure described
in the
Method Selection section (see page 12).
• Fill the cuvette with 10 mL of unreacted sample (up to the
mark) and replace the cap.
• Place the cuvette into the holder and close the lid.
• Press the Zero key. The display will show “-0.0-” when the meter is zeroed and ready for
measurement.
• Remove the cuvette.
• Add 10 drops of HI 937O6A-0 Molybdate reagent.
10 mL