Phosphorus LR

• Add the content of one packet of HI 93706B-0 Phosphorus

Reagent B (Amino Acid) to the cuvette. Replace the cap
and shake gently until completely dissolved.

• Reinsert the cuvette into the instrument.

• Press Timer and the display will show the countdown prior to the measurement or, alternatively, wait

for 5 minutes and press Read. When the timer ends the meter will perform the reading. The instrument
displays the results in mg/L of phosphorus (P).

• Press

or

to access the second level functions.

• Press the Chem Frm key to convert the result in mg/L of phosphate (PO

4

3

¯) and phosphorus pentoxide

(P

2

O

5

).

• Press

or

to return to the measurement screen.

INTERFERENCES

Interference may be caused by:
Sulfide
Chloride above 150000 mg/L
Calcium above 10000 mg/L as CaCO

3

Magnesium above 40000 mg/L as CaCO

3

Ferrous iron above 100 mg/L