Hanna Instruments HI 83225 User Manual
Page 45
45
Potassium MR
POTASSIUM MEDIUM RANGE
SPECIFICATIONS
Range
10 to 100 mg/L
Resolution
2.5 mg/L
Accuracy
±15 mg/L ±7% of reading at 25 °C
Typical EMC Dev. ±2.5 mg/L
Light Source
Tungsten lamp with narrow band interference filter @ 610 nm
Method
Adaptation of the Turbidimetric Tetraphenylborate method. The reaction between
Potassium and reagents causes turbidity in the sample.
REQUIRED REAGENTS
Code
Description
Quantity
HI 93750A-0
Potassium Reagent
6 drops
HI 93750B-0
Powder Reagent
1 packet
REAGENT SETS
HI 93750-01 Reagents for 100 tests
HI 93750-03 Reagents for 300 tests
For other accessories see page 57.
MEASUREMENT PROCEDURE
Note: for sample preparation follow the NUTRIENTS SOLUTION (MR) procedure at page 21.
• Select the
Potassium MR method using the procedure described in the Method Selection section (see
page 12).
• Fill the cuvette with 10 mL of sample, up to the mark.
• Add six drops of HI 93750A-0, replace the cap and swirl the
solution.
• Place the cuvette into the holder and close the lid.
10 mL
- pH 213 (17 pages)
- pH 211R (13 pages)
- pH 21 (9 pages)
- pH 209V (9 pages)
- pH 209R (13 pages)
- HI 99181 (2 pages)
- HI 99171 (2 pages)
- HI 99163N (2 pages)
- HI 99161N (2 pages)
- HI 99141 (2 pages)
- HI 99131 (2 pages)
- HI 99121 (9 pages)
- HI 99104 (2 pages)
- HI 991003 (2 pages)
- HI 991002 (2 pages)
- HI 991001 (2 pages)
- HI 98402 (9 pages)
- HI 98401 (8 pages)
- HI 98308 (2 pages)
- HI 98240 (21 pages)
- HI 98203 (2 pages)
- HI 98202 (2 pages)
- HI 98201 (2 pages)
- HI 98185 (72 pages)
- HI 98172 (52 pages)
- HI 98160N (48 pages)
- HI 98160 (48 pages)
- HI 98150 (17 pages)
- HI 9815 (2 pages)
- HI 981410 (2 pages)
- HI 981408 (2 pages)
- HI 98121 (2 pages)
- HI 98120 (2 pages)
- HI 98113 (2 pages)
- HI 98110 (2 pages)
- HI 98108 (2 pages)
- HI 98107 (2 pages)
- HI 98106 (2 pages)
- HI 98103 (2 pages)
- HI 96108 (2 pages)
- HI 96107 (2 pages)
- HI 96106 (2 pages)
- HI 96104 (28 pages)
- HI 96101 (36 pages)
- HI 9321 (36 pages)