beautypg.com

Hanna Instruments HI 83225 User Manual

Page 44

background image

44

Potassium HR

POTASSIUM ULTRA HIGH RANGE

For samples containing more than 200 ppm of Potassium: follow the sample preparation procedure
described at page 22 for NUTRIENTS SOLUTION (HR). Then add to the graduated cylinder 20 mL of the
prepared sample (for HR) and fill the cylinder with demineralized water from the Demineralizer Bottle up
to the 100 mL mark.
Follow the MEASUREMENT PROCEDURE at page 42.
Read the result in mg/L of Potassium on the display and multiply the reading by 5 to obtain the actual
concentration of Potassium.

INTERFERENCES

Interferences may be caused by:
Ammonium above 10 ppm
Calcium above 10000 ppm as CaCO

3

Chloride above 12000 ppm
Magnesium above 8000 ppm as CaCO

3

Sodium above 8000 ppm