Hanna Instruments HI 83225 User Manual
Page 28

28
Ammonia LR
SPECIFICATIONS
Range
0.0 to 10.0 mg/L
Resolution
0.01 mg/L
Accuracy
±0.1 mg/L ±4% of reading at 25 °C
Typical EMC Dev. ±0.1 mg/L
Light Source
Tungsten lamp with narrow band interference filter @ 420 nm
Method
Adaptation of the
ASTM Manual of Water and Environmental Technology, D1426-92,
Nessler method. The reaction between ammonia and reagents causes a yellow tint in
the sample.
REQUIRED REAGENTS
Code
Description
Quantity
HI 93715A-0
First Reagent
4 drops (6 drops for seawater)
HI 93715B-0
Second Reagent 4 drops (10 drops for seawater)
REAGENT SETS
HI 93715-01 Reagents for 100 tests
HI 93715-03 Reagents for 300 tests
For other accessories see page 57.
MEASUREMENT PROCEDURE
ATTENTION:
Note: for sample preparation follow the IRRIGATION WATER (LR) procedure at page 20.
• Select the
Ammonia LR method using the procedure described in the
Method Selection section (see page 12).
• Fill the cuvette with 10 mL of unreacted sample (up to the mark) and
replace the cap.
• Place the cuvette into the holder and close the lid.
• Press the Zero key. The display will show “-0.0-” when the meter is zeroed and ready for
measurement.
AMMONIA LOW RANGE
10 mL