beautypg.com

Hanna Instruments HI 83225 User Manual

Page 24

background image

24

Ammonia HR

SPECIFICATIONS

Range

0 to 100 mg/L

Resolution

1 mg/L

Accuracy

±1 mg/L ±4% of reading at 25 °C

Typical EMC Dev. ±1 mg/L
Light Source

Tungsten lamp with narrow band interference filter @ 420 nm

Method

Adaptation of the

ASTM Manual of Water and Environmental Technology, D1426-92,

Nessler method. The reaction between ammonia and reagents causes a yellow tint in
the sample.

REQUIRED REAGENTS

Code

Description

Quantity

HI 93715A-0

First Reagent

4 drops (6 drops for seawater)

HI 93715B-0

Second Reagent 4 drops (10 drops for seawater)

REAGENT SETS

HI 93715-01 Reagents for 100 tests
HI 93715-03 Reagents for 300 tests
For other accessories see page 57.

MEASUREMENT PROCEDURE

ATTENTION:
Note: for sample preparation follow the NUTRIENTS SOLUTION (HR) procedure at page 22.

• Select the

Ammonia HR method using the procedure

described in the

Method Selection section (see page 12).

• Fill the cuvette with 10 mL of unreacted sample (up to

the mark) and replace the cap.

• Place the cuvette into the holder and close the lid.

• Press the Zero key. The display will show “-0.0-” when the meter is zeroed and ready for

measurement.

AMMONIA HIGH RANGE

10 mL