beautypg.com

Specifications – Hanna Instruments HI 84430 User Manual

Page 4

background image

7

6

REQUIRED REAGENTS
Code

Description

Quantity/Test

HI 84430 - 50

Titrant (Low Range)

1 mL

HI 84430 - 51

Titrant (High Range)

1 mL

HI 84430 - 55M

Pump Calibration Solution

2 mL for low range
20 mL for high range

Electrode

HI 1131B (included)

Temperature Probe HI 7662-M (included)
Environment

0 to 50 °C (32 to 122 °F); max 95% RH non-condensing

Power supply

12 Vdc power adapter

Dimensions

208 Ч 214 Ч 163 mm (8.2Ч8.4Ч6.4”) (with beaker)

Weight

2200 g (77 oz.)

SPECIFICATIONS

Titrator

Range

Titratable acidity (low range)
mg/L

: 15.0 - 500.0 mg/L as CaCO

3

meq/L : 0.3 - 10.0 meq/L

as CaCO

3

Titratable acidity (high range)
mg/L

: 400 - 4000 mg/L as CaCO

3

meq/L : 8 - 80 meq/L

as CaCO

3

Resolution

Titratable acidity (low range):
0.1 mg/L
0.1 meq/L
Titratable acidity (high range):
1mg/L
1meq/L

Accuracy

5% of reading

Titration method Acid-base titration (total acidity / strong acidity)
Principle

End point titration : 8.3 pH / 3.7 pH

Pump debit

0.5 mL/min

Stirring speed

600 rpm

Log data

Up to 50 samples

pH meter

pH meter

-2.0 to 16.0 pH / -2.00 to 16.00 pH

pH Resolution:

0.1 pH / 0.01 pH

pH Accuracy:

± 0.01 pH

pH Calibration:

1, 2 or 3 calibration points;

3 available buffers (1.68; 4.01; 8.30)

Temperature

manual or automatic from

compensation:

-20 to 120 °C (-4 to 248 °F)

Log data

Up to 50 samples

Temperature

Range

-20.0 to 120.0 °C (-4.0 to 248.0 °F)

Resolution

0.1 °C

Accuracy

±0.4 °C without probe error