TITRATOR GLP INFORMATION
Press MENU while in Titration mode and then GLP.

From this screen it is possible to select between viewing the electrode GLP or the pump GLP.
Press GLP elec. to view the electrode’s last calibration parameters and date.
Press GLP pump to view the pump’s last calibration time and date.
If GLP elec. is pressed one of the next screens will be displayed.

GLP contains a set of information regarding electrode calibration. The following items are included

in electrode GLP: the time and date of the last calibration, offset, slope, electrode condition,

calibration timeout and the calibration buffers. The buffers displayed in video inverse mode are from

the previous calibration.
If a calibration hasn’t been performed the message “Not Calibrated” will be displayed.

If GLP pump is pressed, one of the next screens is displayed.

Press Delete to enter record deleting mode.
Press Delete All to enter all records deleting mode.
Press More to view more information.

Press Unit to convert the result to other unit.
Use the ARROW keys when is displayed to scroll between the log records.
Press ESC to return to the previous screen.
If Delete was pressed the instrument will ask for confirmation.

Use the ARROW keys to focus on the record to be deleted.
Press Confirm to delete the record or ESC to return to the previous screen.
Deleting a record will renumber the list of records.
If Delete All was pressed the instrument will ask for confirmation.

Press Confirm to delete all the records or ESC to return to the previous screen.

If the titrator log is empty the message “No Records!” will be displayed.