beautypg.com

Specifications – Hanna Instruments HI 84432 User Manual

Page 6

background image

6

SPECIFICATIONS

Titrator

Range

Titratable acidity (low range): 15 mL sample

g/100 mL

as citric acid: 0.20 - 1.20% CA

g/100 mL

as tartaric acid

: 0.23 - 1.41% TA

g/100 mL

as malic acid

: 0.21 - 1.26% MA

Titratable acidity (high range): 2 mL sample

g/100 mL

as citric acid: 0.80 - 8.00% CA

g/100 mL

as tartaric acid

: 0.94 - 9.30% TA

g/100 mL

as malic acid

: 0.84 - 8.30% MA

Resolution

0.01%

Accuracy

5% of reading or ± 0.02

Titration method Acid-base titration
Principle

End point titration: 8.20 pH

Pump debit

0.5 mL/min

Stirring speed

600 rpm

Log data

Up to 50 samples

pH meter

pH meter

-2.0 to 16.0 pH / -2.00 to 16.00 pH

pH Resolution

0.1 pH / 0.01 pH

pH Accuracy

± 0.01 pH

pH Calibration

1, 2 or 3 calibration points;

3 available buffers (4.01; 7.01; 8.20)

Temperature

manual or automatic from

compensation

-20 to 120 °C (-4 to 248 °F)

Log data

Up to 50 samples

Temperature

Range

-20.0 to 120.0 °C (-4.0 to 248.0 °F)

Resolution

0.1 °C

Accuracy

±0.4 °C without probe error

Electrode

HI 1131B (included)

Temperature Probe

HI 7662-M (included)

Environment

0 to 50 °C (32 to 122 °F); max 95% RH non-condensing

Power supply

12 Vdc power adapter

Dimensions

208 Ч 214 Ч 163 mm (8.2Ч8.4Ч6.4”) (with beaker)

Weight

2200 g (77 oz.)