Press HELP to view the contextual help, every time you need supplementary information. The

help is customized for every situation that can appear during instrument usage.
Press STIR to start/stop the stirrer.
Press Titrator to enter titrator mode.
Press CAL to access the calibration menu.
Press LOG to memorize the current reading. A message indicating the free log space will be

displayed for a few seconds.

In order to take pH measurements follow the next steps:
• Submerge the pH bulb 2 cm (0,8’’) and the temperature probe into the sample to be tested

and stir gently. Allow time for the electrode to stabilize. When the reading becomes stable

the (unstable measurement) symbol will disappear.

• If the pH reading is less than -2.00 pH or greater than 16.00 pH the closest full-scale value

(-2.00 pH or 16.00 pH) will be displayed blinking.

If measurements are taken successively in different samples, it is recommended to rinse the

electrodes thoroughly with deionized water or tap water and then with some of the next sample

to prevent cross-contamination.
The pH measurement is affected by temperature. In order to have accurate pH measurements,

the temperature effect must be compensated for. To use the Automatic Temperature Compensation

(ATC) feature, connect and submerge the HI 7662-M temperature probe into the sample in the

dedicated guide tube and wait for a few seconds. The “ATC” message will be shown on the

display. Automatic Temperature Compensation will provide pH corrected values at the temperature