• If the computed slope isn’t within the preset limits the meter will display the

message “Wrong Slope”. If the slope is too high the symbol

will be

displayed. If the slope is too low the symbol

will be displayed.

• If the “Wrong Old Slope” error message is displayed, an inconsistency exists

between the current and the previous (old) calibration. Clear the calibration parameters

by pressing Clear and proceed with calibration from the current calibration point. The

instrument will keep all the confirmed values during the current calibration point.

• If the temperature reading is out of the defined temperature range of the buffer (0 to

45°C) the “Wrong Buffer Temperature” error message will be displayed, and the

symbol °C will blink on the display. Calibration cannot be confirmed in this

situation.